Introduction and Aims: Diagnostic Radiology has the largest share in the artificial ionizing radiation sources in the society. One of the most important factors in reducing radiation dose is quality control and ensuring the quality and performance of radiology
Materials and Methods: Ten normal X-ray machines in 5 hospitals affiliated to Kurdistan university of Medical Sciences were studied. Then, 10 standard of quality control
: In some devices the overall filtration parameters’ error and voltage accuracy was higher than acceptable range and needed to be reformed. Also, in some devices the percentage of of radiation field compliance and light field was more than acceptable range based on the existing standards. Some devices in the province had boundary and threshold error.
Conclusion: According to the recommendations of the Commission on Radiation Protection and Medical Physics Society of America, most of the parameters of the devices in question were in agreement with existing standards and were in acceptable range